Test Procedure (ZVM & ZVK) ZVx
1127.8500.60 5.14 E-2
5.2.3.2 Matching Reference Channel Inputs R1 and R2 Channel IN
Test equipment: Calibration Kit ZV-Z32, pair of Test Cables ZV-Z14 (ZVM);
Calibration Kit ZV-Z34, pair of Test Cables ZV-Z15 (ZVK)
Test setup: Description for R2 Channel IN (for R1 Channel IN analogously)
Connect test cable to PORT1.
Settings on network analyzer:
PRESET
AVG IF BW 100 Hz
Record one-port calibration (OSM) at PORT1 over the whole frequency range.
Connect test cables (incl. adapter) with R2 Channel IN.
MARKER Measurement frequency
MEAS S11 (S22)
Measurement:
½ Set measurement frequencies according to test record, measure S11 (or S22 for
R1 Channel IN).
½ Record match values