8-13
Temperature Uniformity Test
Use this procedure to test the temperature uniformity of the
sample block in the GeneAmp PCR System 9600.
Equipment Required
You will need the following equipment:
Temperature Verification System
Testing the Temperature Uniformity
Perform the procedure as described in the following steps:
1
If the heated cover is in the forward position, turn the cover knob
completely counterclockwise, then slide the cover back.
2
Coat all the wells in sample block rows A, C, E, and H with mineral
oil using a cotton swab.
3
Place the probe tray on the sample block with the notch facing the
front of the instrument.
The RTD probe assembly consists of two cones, one of which
measures the temperature of the sample well. The wire is attached
to the cone that does not measure the temperature of the sample
well. This cone is a dummy probe.
4
Place the probe assembly into wells A1 and A2 with the dummy
probe in A2. Thread the probe wire through the notch in the probe
tray (see Figure 8-5). Make sure the probe is connected to the
digital thermometer.