Endosafe
®
- Portable Test System User’s Guide
Copyright 2003 Page C3
PIPTS101-10
INHIBITION/ENHANCEMENT TEST RESULTS
1. When the test is complete, the PTS Reader gives an audible notification that
the assay is finished.
2. At the conclusion of the test, the assay acceptance criteria are displayed on the
screen.
3. The PTS Reader display alternates between the following results:
x Spike #1 Recovery
x Spike #2 Recovery
x Spike #3 Recovery
x Spike #4 Recovery
The PTS Reader continues to display the assay results until cartridge is removed. Once
the assay is complete and the results are noted, remove the cartridge promptly from the
PTS Reader.