Section 2: Sensor and Sampling Techniques32. Sensor and Sampling Techniques2.1 Precautions using the sensorThe Xentaur HTF™ Al2O3 sensor is designed a

Strona 3

XDT User’s Manual42.2 Sensor Technical SpecificationsType: ...Hyper Thin Film HTF™ high capacitance Al2O3.Dewpoint rang

Strona 4 - Warning Labels

Section 2: Sensor and Sampling Techniques52.3.1 In-situ InstallationIn-situ installation is recommended only for measurements where the gas pressure i

Strona 5

XDT User’s Manual62.3.2 Extractive InstallationFor extractive installations we recommend our sample system ESS, which may be equipped with a variety o

Strona 6 - Table of Contents

Section 2: Sensor and Sampling Techniques7inside these materials.• Try to run pipes to the sensor upwards, so that contaminants tend to fall back into

Strona 7

XDT User’s Manual82.5 Troubleshooting unexpected readingsIf erroneous readings are suspected on a newly acquired instrument, compare the serial number

Strona 8 - 1.0 Overview of the XDT

Section 3: Instrument93. Instrument3.1 Precautions using the XDTThe XDT uses state-of-the-art microelectronics to provide a compact full functioning i

Strona 9

XDT User’s Manual103.2 Instrument Technical SpecificationsEnclosure: ...XDT-OEM bare boardXDT-NEMA tabletop or surface mount, wa

Strona 10

Section 3: Instrument113.3 Installation3.3.1 Instrument Mechanical InstallationThe XDT is available in several different physical configurations. Plea

Strona 11 - XDT User’s Manual

XDT User’s Manual123.3.2 Electrical ConnectionsAll connections are made via a 21 contact, pluggable screw terminal block referred to as P1. Cable acce

Strona 12 - NOT RECOMMENDED

XDT User’s ManualiWhen calling your representative for technical support, please have your serial numbers available.The Sensor and Instrument Serial N

Strona 13

Section 3: Instrument13fast acting fuse. Please follow the appropriate installation procedures in the paragraphs below according to the electrical pow

Strona 14 - 2.4 Mechanical Installation

XDT User’s Manual14dent relays corresponding to the HI and LO alarms. The relay contacts are rated at 10A 250VAC or 30VDC, instruments for use in divi

Strona 15

Section 3: Instrument153.3.2.5 Interfacing to the RS-232 optionThe optional RS-232C interface is provided on the P1 connector. The configuration is 96

Strona 16 - 3.1 Precautions using the XDT

XDT User’s Manual163.4.2 Display Conventions1. To display characters with the 7 segment numeric display, the following pseudo-alphanumer-ics are used:

Strona 17

Section 3: Instrument17switches described above.• With a front face plate which has labeled holes placed immediately above the circuit board switches,

Strona 18 - 3.3 Installation

XDT User’s Manual18parts per million by volume computed at the sensor pressure (more about pressure later). The LBS and G/M3 are density measurements,

Strona 19

Section 3: Instrument193.4.4.2 AlarmsThere are two independent optional alarms, they are named HI and LO alarms. Each alarm can activate a single pole

Strona 20 - Section 3: Instrument

XDT User’s Manual205. Push the MODE button. The display will show:the ‘DEWPOINT’ indicator means that we are changing the alarm trigger dewpoint6. Pus

Strona 21

Section 3: Instrument213.4.4.3.1 SpanCheck™ ModeThe instrument is calibrated at the factory with the sensor it is shipped with and does not need to be

Strona 22 - 3.4 Operating the Instrument

XDT User’s Manual22The instrument will retain the calibration even if the power is turned off.Under certain conditions, an over (super) saturated sens

Strona 23

XDT User’s ManualiiXentaur reserves the right to change or modify the product specification and / or appearance at any time without notice. Therefore,

Strona 24

Section 3: Instrument23point the waiting time may have to be increased - specially if the sensor was wet before the procedure is initiated. Consult wi

Strona 25

XDT User’s Manual24DEWPOINT legends will also appear. Use the UP and DOWN buttons to modify this dew-point to match the dewpoint of the reference gas.

Strona 26

Section 3: Instrument25• The instrument will display the measurement using the newly acquired calibration values.The instrument computes the dewpoint

Strona 27

XDT User’s Manual26CAUTION:2), 3), 4) and 5) are set at the factory and need only be modified when a sensor is changed.2. Selecting the sensor type: I

Strona 28

Section 3: Instrument27a. Go to the CAL/ADJ mode and press the UP button, the display will show the current value, in the range of -1.99 to +1.99 foll

Strona 29

XDT User’s Manual28ber can be corrected by reversing the use of the UP and DOWN keys), with the “12” flashing. Pressing the “MODE” button will flash t

Strona 30

Section 3: Instrument29• This configuration is also equipped with a NEMA sealed audio indicator (>80dB @ 3ft) mounted in the instrument enclosure.

Strona 31

XDT User’s Manual30Legend:  denotes a beeping accompaniment to the message./ denotes alternately flashing messages.DISPLAY EXPLANATION REQUIRED A

Strona 32

Section 3: Instrument313.7 MaintenanceThe XDT-PM and XDT-NEMA instrument enclosures may be cleaned when necessary using a damp (moist) cloth.The XDT-O

Strona 33

XDT User’s Manual32

Strona 34

XDT User’s ManualiiiExamine the XDT package for damage or mishandling. If any damage is evident notify the carrier and request an inspection.Unpack th

Strona 35

Glossary33Glossary0/24 mA . . . . . . . . . . . . . . . . range of current for the current loop (0mA=minimum reading, 24mA=maximum reading)4/20 mA .

Strona 36

XDT User’s Manual34used for the circuit (see 4/20)dB . . . . . . . . . . . . . . . . . . . . . see - decibelDB25. . . . . . . . . . . . . . . . . . .

Strona 37

Glossary35impedance . . . . . . . . . . . . . . . the ratio of voltage to current at a specified frequency measured in Ohms, a resistor of the same va

Strona 38 - 3.7 Maintenance

XDT User’s Manual36RG58. . . . . . . . . . . . . . . . . . . a type of coaxial cable - the letters/numbers refer to the electrical characteristic of t

Strona 39

XDT User’s Manual Appendices37Appendix A: Flow Diagram of Operating State User Interface˚C DEWPOINTSETSETHIHI/LOSETHISETLOSETLOSETLOSETHISETSETSETSETS

Strona 40 - Glossary

XDT User’s Manual Appendices38Flow Diagram Operating State User Interface (continued) Manual & Scheduled Self Calibration˚C DEWPOINTHI/LOSETSETSET

Strona 41

XDT User’s Manual Appendices39Appendix B: Flow Diagram of Set-Up State User InterfaceHi/LoSETSETSETLoLoLoHiSETSETSETSETSETSET SET SET SETSET SET SETSE

Strona 42

XDT User’s Manual Appendices40Appendix C: Sensor Mechanicalflats for 1 " wrench14/~1.45"14mm x 1.25mm3/4"-16Viton A O-ringfemale BNC

Strona 43

XDT User’s Manual Appendices41Appendix D: Optional Sensor Fittings

Strona 44 - XDT User’s Manual Appendices

XDT User’s Manual Appendices42Appendix E: XDT Circuit Board DimensionsLOW RELAYHI RELAYNCNOCNCNOCCAUTIONRELAYVOLTAGEooGNDGND10VA50/60HZ100-250VACJ.F.M

Strona 45

XDT User’s ManualivWarrantyThis instrument is warranted to be free from defects in workmanship and materials. Liability under this warranty is limited

Strona 46

XDT User’s Manual Appendices43Appendix E (continued): XDT-PM Enclosure Mounting DimensionsStandard mounting (with clip-on compression brackets)Environ

Strona 47 - Appendix C: Sensor Mechanical

XDT User’s Manual Appendices44Appendix E (continued): XDT-NEMA Enclosure Mounting Dimensions

Strona 48

XDT User’s Manual Appendices45Appendix F: Sensor/SpanCheck™ Theory of OperationThe Sensor is constructed as a capacitor whose dielectric consists of p

Strona 49

XDT User’s Manual Appendices46Appendix G: Dewpoint Response time AnalysisManufacturers of dewpoint sensors will make unbelievable claims of speed. But

Strona 50

XDT User’s Manual Appendices47Of course analyzing on a wet-up step will produce much faster results, because all sensors are fast on wet up, since Al2

Strona 51

XDT User’s Manual Appendices48Appendix H: Sample Gas Filter ConsiderationsSample treatment with pre-filtering may be a necessity in many instances. Ho

Strona 52 - 2O3 and the

XDT User’s Manual Appendices49Appendix I: XDT Grounding Considerations1. The sensor ground is isolated from the signal and frame grounds, however it i

Strona 53

XDT User’s Manual Appendices50Appendix J: Analog Output vs. Dewpoint The current output of the XDT, varies with the measured dewpoint. To use the curr

Strona 54

XDT User’s Manual Appendices51-100°C to +20°C settings, because the 4-20mA provides sufficient resolution to measure the out-put better than the speci

Strona 55 - Tested Instrument (˚C)

XDT User’s Manual Appendices52Appendix K: RS-232C Interface ProtocolThe instrument uses a simple protocol to communicate to a host computer through it

Strona 56 - AC Power

XDT User’s ManualvTable of Contents1.0 Overview of the Dewpoint Transmitter model (DT)...12. S

Strona 57

XDT User’s Manual Appendices53The mode argument is optional and may be specified only if an interval was specified. It may be “N” for numeric reportin

Strona 58

XDT User’s Manual Appendices54The printer should have at least 80 character columns, and it is best to use continuous form paper.“?” commandRequests a

Strona 59

XDT User’s Manual Appendices55Appendix L: Procedure for Exchanging XDT SensorsTo exchange sensors one must know the measured low attenuation of the se

Strona 60

XDT User’s Manual Appendices56Appendix M: Uncertainty in LBS & ppmV calculationsUncertainty of LBS & ppmV calculations due to +/-3°C measureme

Strona 61

Appendix M: Return Authorization Number Request Form Xentaur Return Authorization Number Request Form • Xentaur must approve and assign a Return Aut

Strona 63

57Symbols°C ...17°F ...17AAC power voltag

Strona 64

Index58copper tubing ...6copyright ...icorrosive gases ...

Strona 65

59IndexNPL ...1Ooil ...3operating state .

Strona 66

Index60voltage, mains supply ...10Wwarning labels ... iiiwarranty ...

Strona 67

XDT User’s ManualviAppendix D: Optional Sensor Fittings...41Appendix

Strona 68

XENTAUR CORPORATION, 84F HORSEBLOCK ROAD, YAPHANK, NY 11980 U.S.A. TEL (631) 345-3434 FAX (631) 345-5349

Strona 69

Section 1: Introduction11.0 Overview of the XDTThe XDT is a microprocessor based hygrometer, for measuring moisture content in gases in the range from